- Moisture Meters[6]
- Tubs[1]
- Other Measuring & Analysing Instruments[6]
- Speed Measuring Instruments[6]
- Other Electronic Measuring Instruments[6]
- Other Vehicle Tools[2]
- Temperature Instruments[6]
- Other Security & Protection Products[2]
- Gas Analyzers[3]
- Other Analysis Instruments[1]
- Other Optics Instruments[1]
- Voltage Meters[2]
- Frequency Meters[2]
- Contact Person : Mr. Wan Kexu
- Company Name : Shenzhen Sanpo Instrument Co., Ltd.
- Tel : 0086-755-27863686
- Fax : 0086-755-27863682
- Address : Guangdong,Shenzhen,3/F 3 New Building 43 District Baoan Shenzhen 518101,China
- Country/Region : China
- Zip : 518101
Related Product Searches:High-frequency field intensity meter,High Quality EMF Meter,EMF Meter EMF829
General specification
Metrical ways :digit display, measuring with three axes
Characteristic of directivity : isotropic, three axes
Metrical gear: one continuous gear
Resolution displayed: 0.1mV/m,0.1uA/m,0.1uW/m²,mW/m²,uW/cm²
Reaction time: about 1 second (the measurement value is from 0 to 90% )
Refresh rate displayed: about 0.5 second
Display format: LCD, 4 digits
Alarm: buzzer
Units: mV/m,V/m,uA/m,mA/m,uW/m²,mW/m²,uW/cm²
Displayed value: the instantaneous measurement value, the MAX measurement value, average measurement value and the maximum averaging value
Warning function: you can adjust the setting value and you also can open
the function or close it.
Correction factor(CAL): it can be adjusted
Manual memory and reading data content: 99 groups
Dry battery: 4 batteries of 1.5V
Battery life: about 15 hours
Power off automatically: 15 min
The range of the operation temperature: from 0°C to +50°C
The range of the operation humidity: from 25% to 75%RH
The range of the stored temperature: from -10°C to +60°C
The range of the stored humidity: from 0% to 80%RH
Size: 261cm(length)* 73cm(width)*38cm(height)
Weight( include the batteries): about 220g
Accessories: instruction manual, batteries, carrying case
Electrical specification
The following specification all adapt for the following conditions except for some special explanation:
the meter locate the far-field region of the radiation source and the inductor aim at emissive source.
ambient temperature:+23°C±3°C
relative air humidity: 25% to 75%
inductive mode: electric field(E)
frequency range: 50MHz to3.5GHz
measuring range: continuous (CW)signal(f>50MHz): 20mV/m to 108.0V/m;
53uA/m to 286.4mA/m;
1uW/m² to 30.93W/m²;
0uW/cm² to 3.093mW/cm²
dynamic range: about 75dB
absolute error(lower than 1V/m and 50MHz): ±1.0dB
frequency response:
the typical correction factor of the inductor bringing into the considerations: ±1.0dB (50MHz to 1.9GHz)
±2.4dB (1.9GHz to 3.5GHz)
isotropic windage: about ±1.0dB(f>50MHz)overloading limit: 0.42mW/cm2(40V/m)
temperature response: (0 to 50°C): ±0.2dB